MAGICL 6000i
Chemiluminescence Immunoassay Analyzer
Intended Use
MAGICL 6000i Chemiluminescence Immunoassay Analyzer (hereinafter called MAGICL 6000i) automates the detection and quantification of biomarkers for thyroid disease, tumor, fertility, infectious disease, bone metabolism, cardiac disease, diabetes mellitus, inflammation, anemia, and immune globulin by analyzing human serum, plasma, whole blood and urine samples.
Features
Reagent Unit
24 reagent positions, supporting on-board loading
Automatic recognition of reagents by QR code scanning
2-8 °C refrigeration system, with one-month shelf life of regents on board
PE-film puncture technique, preventing reagents evaporation
Sampling Unit
Auto mixing and uncapping of sample tubes
Liquid level detection function, adding sample more accurately
Sample Unit
10 sample racks;50 positions
Video recognition system, acquiring real status of sample liquid level and avoiding inaccurate sampling caused by bubbles
Emergency sample test in priority
Test Unit
Combining washing station, test station and incubation station in one unit
Small size with high throughput
Unique design of washing station, excluding whole blood factors influence
Consumable Unit
Integrated dual consumable module
Continuous consumable change without test pauses for easy operation
Technical Specifications
Test Principle | Acridinium ester direct chemiluminescence |
Sample Positions | 50 |
Sample Type | Serum/Plasma/Urine/Fingertip blood |
Application Scenarios | Laboratory |
Test Mode | Batch/Random/STAT |
Data Transmission | LIS/HIS connectivity |
Sensing Functions |
Liquid level detection, bubble detection, probe clogging detection, temperature monitoring, consumables inventory monitoring |
Test Items